Cannon-instrument miniPV-X Bedienungsanleitung Seite 19

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 130
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 18
13
CANNON
®
miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
The Calibration window
Compare this value to your archive of the calibration constant data (if
available) for bulb 1. The values should be identical.
3. If they are not, place your cursor in the appropriate field, delete the
entry, then type the correct values for the constant in the text box.
4. Click Update tube
.
5. Use the bulb spin controls
to select the other bulb(s) for which
you noted calibration constant errors. Input the correct values for
each.
Make certain to click Update Tube after you have corrected
C and E calibration values for each bulb BEFORE selecting the next
bulb.
6. When you have entered corrected constant values for each bulb on
both tubes, click
to exit the Calibration window.
You have verified the software configuration of VISCPRO
®
. To test
samples with your instrument, follow the instructions in Chapter 2. For
additional details regarding operating procedures for your instrument or
software, consult the appropriate section of this manual.
Checking Wash and Advanced Settings
To check Wash Settings for your miniPV Series instrument, click Config-
ure from the VISCPRO menu options and select the desired instrument.
Then click Tray Settings: Wash. The Tray Settings: Wash window
will appear.
Seitenansicht 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 129 130

Kommentare zu diesen Handbüchern

Keine Kommentare